Overview

  • Clinicians can implement patient therapy sooner than before as testing can be conducted anywhere.
  • i-Smart 300 portable analyzers meet the need for critical care testing in the laboratory, OR, ER, ICU and other hospital location.

Features

01. Maintenance Free

  • Proven quality through the previous Electrolyte Analyzer.
  • Disposable Cartridge contains all sensors, reagents, waste bag, tubing and sample probe.
  • Replacing a cartridge is all you have to do for maintenance!

02. Reliable and Easy to Use

  • Audio guided menu driven touch-screen operation.
  • Self-cleaning sample probe helps to maintain sample integrity.
  • Cutting-edge sensor technology helps to ensure superior quality of measurement.
  • Self-contained reagent cartridge helps to minimize exposure to biohazards.

02. Fast and Flexible

  • RFID allows automatic cartridge information loading.
  • Fast patient results measurement takes only 50 seconds.

  • Fully automated calibration and system performance monitoring.

  • Built-in barcode scanner for entry of patient information, operator identification and QC data.

03. Simple Operation

  • Requires small sample size, ideal for all patient types.
  • Analyzes syringe and capillary samples without adaptors.
  • Small portable and weights less than 7.1 kg.

04. System Connectivity

  • Onboard QC program calculates and stores data.
  • Seamless integration with your LIS/HIS via standard ASTM interface protocol.
Operating parameters Cartridge Instrument Power requirements
Measuring range Choices offering Computer Voltage
pH : 6,500 ~ 8,000
pCO2 : 5.0 ~ 150.0 mmHg
pO2 : 5 ~ 700 mmHg
cNa+ : 80 ~ 200 mmol/L
cK+ : 1.0 ~ 20.0 mmol/L
cCa2+ : 0.25 ~ 5.00 mmol/L
cCl- : 50~150 mmol/L
Hct : 10~70%
100 tests / 2 weeks, 3 weeks
200 tests / 2 weeks, 3 weeks
300 tests / 2 weeks, 3 weeks
1.6 GHz processor
2 GB RAM
8 GB flash storage
100 ~ 240 VAC
Frequency
50/60 Hz(±10%)
Storage condition Operating system Power adaptor
15~25 ℃
Windows® Embedded
standard 7
AC-DC power supply
(Output: DC24V, 2.7A)
Shelf life
6 months from
manufacturing date
Display Battery operation
8 inch TFT LCD Touch Screen
Max. 2 hours (Li ion cells)
Derived parameters Components Printer Interface
pH(T), pCO2(T), pO2(T), ctCO2
cHCO3-, cHCO3-(std), BE(ecf),
BE(B), Anion gap(K), ctHb,
Ca++(7.4), pO2(A-a)
Sensors
Cal 1, Cal 2
Reference solution
Sampler & Waste bag
Valve and tubing
2 inch thermal printer(built-in)
Built-in barcode
scanner
4 USB ports
Serial (RS232) port
Ethernet (RJ45) port
VGA port (for external screen)
Dimensions / Weight
257(W) x 394(H) x 252(D) mm
7.1kg

Sample type Dimensions / Weight
Heparinized whole blood
182(W) x 148(H) x 100(D) mm
1.4 kg
Sample volume
80 ㎕
Sample analysis time
50 seconds
Calibration
Automatic or manual